|
Measuring Electrical Breakdown of a Dielectric-Filled
Trench Used for Electrical Isolation of Semiconductor Devices
By Rishi Gupta, Zyvex Instruments
Download a printable copy here.
Introduction
Semiconductor devices employ insulating dielectric materials
for electrical isolation between active elements and layers
that are susceptible to electrical breakdown. The breakdown
voltage is the level at which the insulating dielectric begins
to allow charge flow. Unlike conducting materials, this charge
flow tends to be non-linear. That means that below a threshold
voltage no charge will flow, and at or above that voltage
a rush of charge will flow. This rush is termed “avalanche
breakdown,” which is a runaway process resulting in
a current spike. Once the charge begins to flow, the dielectric
material properties become unpredictable. To properly characterize
a given film, breakdown voltages must be repeated over different
samples.
Several mechanisms give rise to electron avalanche, one of
which is by impact ionization.1 Impact ionization
occurs when either an internal or external field, acting on
the dielectric accelerates electrons towards the anode, thereby
facilitating collisions with ubiquitous neutrals. The result
is the formation of electron-hole pairs. Each electron-hole
pair gets polarized by the electric field and accelerated.
In doing so, it can ionize more neutrals, even though the
holes are generally considered to be relatively immobile.
This is a continuous process resulting in a rush of free carriers.
Another mechanism is by field-emission breakdown. Field-emission
breakdown occurs when critical field strength causes electrons
to escape from the valence band to the conduction band.2
Conduction electrons are mobile charge carriers and diminish
the insulating properties of the dielectric.
Breakdown voltage is directly related to the dielectric thickness.2
Its counterpart, dielectric strength, indicates the relationship
between thickness and breakdown voltage, and is expressed
as a ratio of voltage to thickness. In thin dielectrics (<3
microns), thermal effects due to local Joule heating can cause
thermal breakdown at sufficiently high temperatures. This
correlates to electrical conductivity through an exponentially
increasing function of the temperature. Other factors that
contribute to breakdown include occluded particles, surface
and material contamination, and water vapor. Each can influence
breakdown; actions must be taken to eliminate their contribution
to breakdown voltage measurements. Humidity, for example,
reduces the resistance of most dielectrics, thus increasing
the return current (the current that opposes a charge build-up).3
Contamination can contribute to leakage currents and charge
mobility across isolation areas. Occluded particles such as
alkalis or halides can act to increase the breakdown strength.2
The S100 Nanomanipulation System (Figure 1)
can function as a nano- and microprobe and is an ideal tool
for dielectric breakdown voltage measurements. When operated
inside an scanning electron microscope (SEM), the vacuum environment
minimizes moisture.

Figure 1. Zyvex S100 Nanomanipulator
can fucntion as a nano- and microprobe, and is an ideal tool
for dielectric breakdown voltage measurement.
The SEM also allows the material to be imaged to ascertain
damaged areas and perform a failure analysis in situ. The
S100 can also operate in ambient environments for sub-micron
probing. The modular design of the system allows for quick
interchangeability between the SEM and the optical microscope.
Such versatility makes the Zyvex S100 the premier system for
manipulation and nanometer-scale/micron-scale device probing.
Device Probing
The S100 System is a manipulator designed for nanometer precise
resolution. It uses two types of mechanisms for motion: coarse
positioners consisting of piezo-motor based, scratch drive
actuators allowing linear translation; and fine positioners
consisting of piezo tube/piezo stack actuation, which are
mounted to the coarse positioners. The positioners are controlled
by a joystick/keypad, and move in each of the cardinal directions
with a resolution of less than 5 nanometers and a range of
12 millimeters. The piezo-electronically actuated center-sample
stage rotates clockwise and counterclockwise with a resolution
of 5 microradians. At the end of each fine positioner is a
five-pin plug into which probes or other end effectors can
be affixed. These probes are affixed into the S100 positioners
by a friction fit. One can check connectivity through the
system with a multi-meter, as all of the probes are electrically
accessible via the S100 cabinet. Positioners are used to move
the probes into place, and the sample is aligned with the
rotational stage.
The S100 may be operated under ambient environments using
an optical docking station or in vacuum under an SEM. Each
has its advantages, and the required environment is a function
of the application. The vacuum environment maintains a low
relative humidity, keeping moisture from precipitating electrical
breakdown. Using a scanning electron microscope will enable
the user to image and probe nanometer-scale devices. However,
vacuum environments require more preparation time. For “quick
and dirty” measurements, one would opt for an in-air
set-up to minimize the amount of time spent on the experiment.
In-air testing can be performed quickly (with short preparation
times with the S100), but data acquired from in-air testing
is susceptible to further scrutiny if variables such as humidity
are not taken into consideration.
One can apply a field and measure current using probe style
end-effectors controlled by the S100 System. In either environment,
to properly gauge breakdown voltage, one must eliminate alternative
paths for current to flow. The current must be a strict measure
of the charge flow caused by the applied field across the
dielectric. If there are alternative paths for charge to flow
between the probes, current will be detected before dielectric
breakdown, and the data will not accurately reflect the breakdown
voltage.
When contacting the surface, simply touching the contact
pads, electrodes, or dielectric surface may not be sufficient
to pass current (due to contact resistance). Contact resistance
occurs at the boundary between two materials. It can be attributed
to many things such as oxidation layers, contamination layers,
points of contact, etc. Oxide can grow on metal probes and
results in poor ohmic contact. Contamination layers can act
as insulation between the metal probe and contact pad. To
ensure contact, apply the probes onto the contact area with
enough force to make them rub against the surface.
Additionally, ensure that the coarse positioning motors are
switched off before taking measurements. The noise from the
motor will interfere with the electrical signals.
Considerations for Probing In Air
Zyvex offers a convenient Optical Docking Station option
that allows the S100 head unit to be used in-air under an
optical microscope. The optical docking station consists
of an X, Y-translation stage with an S100 mounting ring,
a long working distance optical stereo microscope, ring
light, and an adaptor assembly for the cables. All of these
components are mounted neatly onto an optical breadboard
for easy adjustment.
Essentially, one can use the S100 operated in-air as a
joystick controlled micro-prober station. If possible, operate
the S100 in a clean room to eliminate dust, control humidity,
and prevent contaminants from building up on the devices.
High humidity environments should be avoided due to the
influence of excessive water vapor on breakdown voltage.
Likewise, low humidity labs should be avoided due to the
increased potential for electrostatic discharge through
the device.
Considerations for Probing In Vacuum (SEM)
Scanning electron microscopes are typically used to image
metal surfaces or metal coated samples. The electron beam
impacts the target surface, and scattered electrons are
collected by a detector and processed for imaging. The sample
must be conducting and must be held at ground potential.
This is so that the electrons from the e-beam will either
scatter or be dissipated through the SEM. If charging occurs,
the electron detectors will be overloaded. Though this may
not cause any damage to the detectors, imaging will be impossible.
Samples are held to ground by attaching them to metal stubs
with double-sided conducting tape or silver paint. The metal
stubs are affixed into the SEM with set screws or into the
S100 with conducting tape. The S100 is held to ground potential
through the SEM translation stage.
After moving all probes into contact with the device, turn
off the electron beam and the motors. Perform all electrical
measurements with the electron beam and motors off to reduce
noise and charging effects. If repositioning is necessary,
repeat the steps above.
Special Devices
Many devices have exposed dielectric layers, such as microelectromechanical
systems (MEMS). These devices will need special attention
for in-vacuum probing because they are more difficult to
isolate electrically. One way to isolate them is by placing
them on an insulating surface, like a printed circuit board
(PCB).
It may seem that imaging with the SEM would not be possible
since the devices have no discharge path. Typically, placing
a device on a non-conducting surface will result in the
sample charging along with the surface. Upon attempting
to image, one would notice the overall brightness increase
very quickly over time (see
video here). This increasing brightness is the sample
charging. As the S100 probes come into contact with the
device (Figure 2), the brightness will
suddenly decrease, indicating that the probe has dissipated
the charge build up (keep in mind that higher target voltages
result in more charging).

Figure 2. Brightness decreases when
probe contacts the isolated device. See video.
Because the S100 can image and perform breakdown voltage
measurements on free-standing, electrically isolated devices
in an SEM, it is an ideal tool for the application.
Figure 3 shows an example of a four-probe
configuration to measure breakdown of an isolation trench
separating two actuators of a MEMS gripper.

Figure 3. SEM micrograph of the probe
configuration for a four probe IV measurement. The support
structure is a PCB and is charging. The device-under-test
is a thermally actuated MEMS device, and is not charging
because the probes are acting to dissipate the surface electrons.
The bright area on the top of the picture shows the PCB
material charging. The MEMS device would also be charging
similarly if the probes were not making contact to the surface.
The isolation trench, seen in the center of the device,
is a silicon nitride trench (Si3N4)
filled with polycrystalline silicon for low cost, structural
support (Figure 4).

Figure 4. Schematic of the isolation
trench from Figures 2 and 3. The silicon nitride forms the
insulating barrier. The trench is filled in with polysilicon
for low-cost structural reinforcement.
Ensure that the input/output to each probe in use is either
capped or connected to the test station. The technique is
not applicable to high-resolution imaging because charge
effects will begin to occur and will disrupt the operation
of the SEM. The dissipation is attributed to leakage of
current through the S100 wiring and cable assemblies. For
proper high-resolution imaging, a true ground plane is required.
Also, keep the device-under-test as far from the external
pole of the SEM as possible to avoid accidental discharge
and damage to the microscope.
The video
shows SEM imaging of the isolation trench of a thermally-
actuated MEMS gripper. The gripper requires electrical and
thermal isolation for the silicon beams to bend to the correct
displacement per unit voltage. Here, the isolation trench
is being tested for electrical breakdown. The trench is
a silicon nitride lined trench that is filled with polycrystalline
silicon (Figure 4). Once the probe makes contact to the
surface of the device, the charge gets dissipated. This
is mainly due to leakage currents through the S100.
Measuring Breakdown Voltage
The best way to characterize a dielectric for electrical breakdown
is to perform a voltage sweep across it. Both types of electrical
breakdown (impact ionization and field-emission breakdown)
are field dependent. That is, there is a critical field strength
at which electrons in the dielectric become influenced by
the external field. In order to ascertain this critical field
strength, one plots current versus voltage as the voltage
is swept through a predetermined range.
The range of the voltage sweep depends on the dielectric,
and the user must determine this value. Keep in mind that,
along with a range, one must consider a sweep rate. Dielectrics
are susceptible to local heating effects that cause thermal
and electrical breakdown, particularly in thin dielectrics.
The resulting plot should start as an open circuit followed
by a sharp current spike at the breakdown voltage (see Figure
5 for a representative plot).

Figure 5. Breakdown of a 50 micron
by 2 micron polysilicon-filled nitride isolation trench on
a silicon-based MEMS device.
The current will continue to spike until it reaches a threshold
which is set either by the protection circuitry or by the
hardware. Allowing the hardware to limit the current could
result in damage to the S100. It is best to use the test station
or some external circuitry to limit the current to 1 mA. The
plot displayed here shows a 1 mA protection cut-off enabled
with the test station software (Keithley SCS-4200).
A Word on Safety
Dielectrics are typically designed to withstand a voltage
level of two to three times the operating voltage of the
device. The S100 can be used with voltage levels up to 500
V and currents up to 1 mA. These current and voltage levels
are not arbitrary; they are the maximum values that the
S100 can carry without potential damage to the system.
Conclusions
The S100 is the perfect instrument for performing electrical
characterization measurements on semiconductor devices. Its
versatility is unmatched since it offers the opportunity to
measure breakdown voltages in-air and in-vacuum. Working in
vacuum helps to minimize the possibility of precipitating
a dielectric electrical breakdown by humidity-induced effects.
In addition, imaging in an SEM while taking breakdown measurements
can allow damaged sites to be ascertained as well as dielectric
thicknesses to be measured. Working in air with an optical
docking station allows for shorter preparation times and faster
experiments. Because the S100 can be moved quickly and easily
from vacuum to ambient environments, can provide precision
movement for sub-micron device probing, and can accommodate
up to 500 V at 1 mA, it is the ideal tool for measuring dielectric
breakdown.
References
1. Ben G. Streetman and Sanjay Banerjee, Solid State Electronic
Devices,
New Jersey: Prentice Hall, 2000.
2. John J. O’Dwyer, The Theory of Electrical Conduction
and Breakdown in Solid Dielectrics, Oxford: Clarendon Press,
1973.
3. “ESD and Humidification,” ESD Systems’
ESD Technical Newsletter, Issue 3, Vol 1, September 1998.
back to top
| ©
Copyright 2008, Zyvex Instruments. All Rights Reserved. |
|